      • Nitrous Nitrogen Water Quality Online Analyzer
      Nitrous Nitrogen Water Quality Online Analyzer
      • Applicable Medium
        Water
      • Not applicable medium
        Gas
      • Product Introduction
        Nitrous nitrogen (NO) refers to the nitrogen content in soluble and suspended particles in water, including nitrite nitrogen, nitrate nitrogen, inorganic ammonium salts, dissolved ammonia, and nitrogen in most organic nitrogen-containing compounds.

      Working principle

      Quantitatively determine nitrate by utilizing the absorption of nitrate ions at a wavelength of 220nm. Dissolved organic matter also absorbs at 220nm, while nitrate ions do not absorb at 275nm. Therefore, another measurement was taken at 275nm, and the corrected absorbance A was calculated based on the formula A=A220-2A275. To calibrate nitrate nitrogen values.

      Product Features

      Modular design allows each component to operate independently, improving the efficiency of the instrument;

      ? Various types of signal outputs: printers RS485/RS232、4-20mA、GPRS、 Relays, etc;

      Various sample making methods: manual control, timed (interval and hourly) control, 4-20mA current input control, switch control, and other forms;

      Automatic compensation for measuring light source attenuation, good stability, and small drift;

      Wide range of measurement, fast sampling speed, and high accuracy;

      Low dosage of medication, maintenance free for one month;

      Optoelectronic non-contact measurement, high measurement accuracy, and high operational reliability;

      Adopting high-resolution industrial grade color touch screen, user-friendly interface, configuration design, and simple operation.


      application area 

      Online monitoring of industrial pollution source wastewater

      Online monitoring of industrial process water usage

      Online monitoring of water quality at the inlet and outlet of municipal sewage treatment plants

      Online monitoring of river, lake, reservoir, and groundwater quality

      Technical Parameter

      Testing principle: UV spectrophotometry

      Measurement range: 0-100mg/L (expandable)

      Colorimetric wavelengths: 220nm and 275nm

      Minimum cycle: 40min

      Accuracy: ± 10%

      Repetitive error: ± 10%

      Zero drift: ± 5%

      Range drift: ± 10%

      Analog output: 1 4-20mA output (optional 2 channels), maximum load resistance of 750 Ω

      Digital output: RS232/RS485

      Switch output: 1 relay output (optional 2)

      Other outputs: Printer output or sending SMS to the target phone number via GPRS to create samples

      Data storage: It can store measurement data for three years and can be stored cyclically

      Data export: Measurement values can be imported into a USB drive through a USB port and saved

      Power supply: AC220 ± 10% V, 50 ± 10% Hz

      Dimensions: Height 1430 x Width 500 x Depth 403 (mm)